| Model | Cpap |
| Pressure Range (cm H2O) | 3 - 10 |
| Portable | Yes |
Minimum order quantity: 1 Piece
• CPAP range (mean): 3 – 10 cmH₂O
• Operational input flow range is 4 – 15 L/min.
• An initial range of 6 – 8 L/min is recommended.
• Gas Leakage @ 60 cmH₂O: <30 mL/min
• Humidification Output (Invasive Mode)
>33 mg/L from 4 to 15L/min over temp range 20 °C to 26 °C
• Resistance to flow @ 2.5 L/min Inspiratory Limb: 0.09 ± 0.032 cmH₂O
Expiratory Limb: 0.05 ± 0.022 cmH₂O (including 0.02 cmH₂O measurement uncertainty)
• Compliance @ 60 cmH₂O:
0.8 ± 0.16 mL/cmH₂O
(including 0.07 mL/cmH₂O measurement uncertainty).
• Tube minimum internal diameter: Inspiratory: 10.9 mm
Expiratory: 10.0 mm
₹ 160000 / Piece Get Latest Price
| Size | Medium |
| Model | Bubble cpap with comressor |
| Pressure Range (cm H2O) | 3 - 10 |
Minimum order quantity: 1 Piece
• CPAP range (mean): 3 – 10 cmH₂O
• Operational input flow range is 4 – 15 L/min.
• An initial range of 6 – 8 L/min is recommended.
• Gas Leakage @ 60 cmH₂O: <30 mL/min
• Humidification Output (Invasive Mode)
>33 mg/L from 4 to 15L/min over temp range 20 °C to 26 °C
• Resistance to flow @ 2.5 L/min Inspiratory Limb: 0.09 ± 0.032 cmH₂O
Expiratory Limb: 0.05 ± 0.022 cmH₂O (including 0.02 cmH₂O measurement uncertainty)
• Compliance @ 60 cmH₂O:
0.8 ± 0.16 mL/cmH₂O
(including 0.07 mL/cmH₂O measurement uncertainty).
• Tube minimum internal diameter: Inspiratory: 10.9 mm
Expiratory: 10.0 mm